Hanna Instruments HI-93755-01 Alkalinity Liquid Reagent, 100 tests

Sale price£34.67 Regular price£36.50
Save £1.83

Alkalinity reagents for use with photometers within the range of 0 to 500 mg/L. Bottle comes with a 1ml syringe. This reagent provides enough to do 100 tests.

The HI-93755-01 Alkalinity liquid reagent is suitable for use with Hanna photometers.

This reagent provides enough to do 100 tests.

Range: 0 to 500 mg/L.

MEASUREMENT PROCEDURE

  • Select the Alkalinity method using the procedure described in the METHOD SELECTION section.

Note: If tutorial mode is disabled, follow the measurement procedure below. If the tutorial mode is enabled, press Measure and follow the messages on the screen.

  • Fill the cuvette with 10 mL of unreacted sample (up to the mark). Replace the plastic stopper and the cap.
  • Insert the cuvette into the holder and ensure that the notch on the cap is positioned securely in the groove.
  • Press Zero. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
  • Remove the cuvette. 10 mL

METHOD PROCEDURE

  • Use a 1 mL syringe and add 1mL of HI-93755 Alkalinity Reagent to the sample.
  • Replace the plastic stopper and the cap. Invert 5 times to mix.
  • Insert the cuvette into the holder and ensure that the notch on the cap is positioned securely in the groove.
  • Press Read to start reading. The instrument displays the results in mg/L of calcium carbonate (CaCO3).

Interference may be caused by:

  • Chlorine must be absent, to remove the interference add one drop of HI-93755-53 Chlorine Removal Reagent to the unreacted sample.

Frequently Asked Questions

Clear answers to help you choose the right water testing solution with confidence