Trace2o HydroTest® COD System HTCOD Trace2o
The HydroTest COD System HTCOD is designed for the determination of chemical oxygen demand index (COD) using the small-scale sealed-tube method.
The industry standard dichromate method can be considered as an estimate of theoretical oxygen demand. With a wide measuring range from 0 – 15,000 mg/L O2, the HTCOD is ideal for waste water testing, offering a small handheld photometer with two LED light sources housed inside a waterproof sample chamber.
The highly-sensitive equipment is compatible with mercury-free tube tests, with optional chloride compensation to remove interference, and utilises the popular photometric determination method.
Standard solutions can be used to verify operation and accuracy of the instrument.
The thermoreactor is compatible with other products in the HydroTest range, and can also be utilised for tube test digestion of TOC, total Chromium total Nitrogen and total Phosphate.
The thermoreactor operates at three pre-set temperatures, with three pre-set reaction times, and offers automatic switch-off functionality with audible alarm.
Features & benefits of HTCOD
Feature | Benefit |
Small-scaled sealed-tube dichromate analysis method | Industry standard mercury-free testing compliant with latest environmental protection legislation |
Extensive detection range | Lower ranges (0 – 150, 0 – 1,500) are perfectly suited to drinking water applications, whilst the higher range (0 – 15,000) is ideal for waste water testing |
Flexible thermoreactor design with safety screen | Suitable for a full range of test tube digestion methods, fully compatible with health and safety requirements |
Fast, effective analysis | Pre-measured tube tests contain requisite reagent – just add water sample, heat in thermoreactor, and measure with photometer |
Waterproof, chemically resistant photometer housing | Perfect for use in a busy laboratory environment, with protection from accidental chemical spills |
Technical Specifications
Ranges
COD (LR) | 0 – 150 mg/L O2 |
COD (MR) | 0 – 1,500 mg/L O2 |
COD (HR) | 0 – 15,000 mg/L O2 |
Technical specifications (Photometer):
User Interface | 4 digit backlit LCD, 4 button rubber keypad |
Operation | Analytical principle: Single parameter fixed wavelength visible absorption photometer Wavelength accuracy: ± 1 nm Dual LED light source with long term stability Wavelengths: λ1 = 610 nm, λ2 = 430 nm Accuracy: 3% FS @ 20 – 25°C Resolutions: 0.01Abs Power supply: 4 x AAA/LR6 alkaline cells, provides 17 hours operation, > 5000 tests |
Environmental | Operating temperature: 5 - 40°C Storage temperature: 0 - 70°C Humidity: 30 – 90% (Non-condensing) Chemically-resistant housing |
Data | 16 data sets Data download via IR link (required T2o PC-Link module) |
Approvals | CE conformity; IP68 water resistance |
Dimensions | TBA |
Weight | TBA |
Technical specifications (Thermoreactor):
Selectable temp | 100 / 120 / 150 °C |
Temperature stability | ± 1 °C at probe |
Selectable time | 30 / 60 / 120 / min. and continuous operation (∞) |
Heating up time | From 20°C to 150°C in 12 min. |
Power supply | 230 V / 50-60 Hz or 115 V / 50-60 Hz (switchable) |
Power | 550 W |
Probe type | Pt100 A class |
Regulation | Microprocessor |
Overheating protection | At the aluminium block at 190 °C |
Beeper | Max. 88 dB (piezo buzzer) |
Aluminium block | 24 holes, 16.2 mm ± 0.2 mm |
User interface | Splashproof membrane keypad, LED indicators |
Materials | Housing - ABS Protection grid - PPS Lid - PC Block insert - PBT Heating block - Aluminium |
Dimensions | 248 x 219 x 171 mm |
Weight | 3.9 Kg |
Environmental conditions | 10 – 40 °C Max. 85 % rel. humidity |
Approvals | CE conformity |

Frequently Asked Questions
Clear answers to help you choose the right water testing solution with confidence
Which brands of water testing kits do you supply?
Do you provide kits for both home and professional use?
Yes — whether you’re a pool or spa owner, aquarium keeper, utility company, or a university laboratory, we have the right solution for your needs.
Can I arrange servicing or calibration for my photometer?
Absolutely. We also provide servicing and calibration to keep your photometers accurate and reliable.
How quickly do you dispatch orders?
Orders placed before 2:30 pm (Monday–Friday) are dispatched the same day. We also ship on Saturdays for your convenience.
Which couriers do you use for delivery?
We deliver through UPS, DPD, and Royal Mail. You can also choose to upgrade to next-day delivery at checkout.
What is your return policy?
We accept returns within 14 days of delivery provided the items are unused, in their original packaging, and in resalable condition. Please contact our team before returning any items.
Can I place bulk or business orders?
Yes, we welcome bulk, trade, and institutional orders. For tailored pricing or large quantities, please get in touch with us directly.
How can I get help if I’m unsure which kit to buy?
You can call us at 📞 +44 (0)115888 2726 and our team will guide you in choosing the most suitable kit for your needs.